SYSTEM MESSAGE
Battery Low
The battery is almost used up. Connect the AC power supply in order to continue
operating the machine.
(3) When the device is printing EKG results, there is a lot of noise added to the signal:
Check whether the AC power supply filter is set up properly. If it is still noisy, even
though the AC power supply filter is ON, connect the ground wire to the GND electrode
of the system. Do not use the GND with that of AC power supply. Instead, connect GND
to the patient’s bed or any other metal around the device that is connected to the
building’s wall.
* If you continue to have problems even after following the aforementioned points, or
other problems that are not mentioned above, contact Medical Econet GmbH’s
customer service.