INSTRUCTIONS MANUAL
-121-
FIG. 36
l
4
1
REPLACE THE NEW UV CARTRIDGE
N.B.: Using UV not recommended by the manufacturer could damage the electronic
valves and will invalidate the warranty.
Procedure:
1. Open the upper door on the right side (ref 4, Fig.36),
2. Press quick connection button (ref 1, Fig.36) to disconnect the UV cartridge.
3. Lift the UV cartridge out of its lodging (ref l, Fig.36),
4. Insert the male fitting of the new UV cartridge into the quick connection and replace the
cartridge in its lodging.
NOTE: Replace the cartridge taking care not to exert pressure on the scale in order not to
damage it.