- 66 -
A2.2 Electromagnetic Immunity
Guidance and manufacturer’s declaration electromagnetic
immunity
H10 is intended for use in the electromagnetic environment
specified below. The customer or the user of the oximeter should
assure that it is used in such an environment.
Immunity
test
IEC/EN
60601 test
level
Complian
ce level
Electromagnetic
environment-
guidance
Electrostat
ic
discharge
(ESD)
IEC/EN61
000-4-2
±8 kV contact
±15 kV air
±8 kV
contact
±15 kV
air
Floors should be
wood, concrete
or ceramic tile.
If floors are
covered with
synthetic
material, the
relative
humidity should
be at least 30%.
Electrical
Fast
Transient/
Burst
IEC/EN61
000-4-4
±2kV for
power supply
lines
±1kV for
input/output
lines (>3m)
N/A
Mains power
quality should
be that of a
typical
commercial or
hospital