A2.4 Recommended Separation
Distances
Recommended separation distances between
portable and mobile RF communications equipment and H10
H10 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of H10
can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications
equipment (transmitters) and H10 as recommended below, according to
the maximum output power of the communications equipment.
Rated
maximum
output power
of
transmitter(W)
Separation distance according to frequency of
transmitter(m)
/
80 MHz to 800 MHz
Pd
2.1
800 MHz to 2.7 GHz
Pd
3.2
0.01
/
0.12
0.23
0.1
/
0.38
0.73
1
/
1.2
2.3
10
/
3.8
7.3
100
/
12
23