SE-12 Series Electrocardiograph User Manual EMC Information
- 101 -
61000-4-3
80 MHz to 2.7 GHz
80 MHz to
2.7 GHz
Pd 3.2
800 MHz to 2.7 GHz
E/6 Pd
at RF wireless
communications equipment bands
(Portable RF communications equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm (12 inches) to
any part of the 12-channel
electrocardiograph, including cables
specified by the manufacturer).
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.