SE-12 Series Electrocardiograph User Manual Operation Preparations
- 31 -
CAUTION
The disposable electrodes can only be used for one time.
Disposable Electrode (clip style): Clip/Snap/Banana Socket Adaptors
Disposable Electrode Connection (Clip Style)
1) Align all lead wires of the patient cable to avoid twisting, and connect the clip/snap/banana socket
adaptors to the patient cable.
2) Clean the electrode areas on the body surface with 75% alcohol.
3) Attach the disposable electrodes to the electrode positions on the body surface.
4) Clip the disposable electrodes with the clip/snap/banana socket adaptors.
Snap/Banana Socket Adapters Disposable Electrode (Snap Style)
Disposable Electrode Connection (Snap Style)
1) Align all lead wires of the patient cable to avoid twisting, and connect Snap/Banana Socket Adapters
to connector of patient cable.
2) Clean the electrode areas on the body surface with 75% alcohol.
3) Attach the disposable electrodes to the electrode positions on the body surface.
4) Connect Snap/Banana Socket Adapters to the disposable electrodes.
3.4 Inspection Before Power-On