SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
- 84 -
disinfection methods.
2. Do not use chloric disinfectant such as chloride, sodium hypochlorite etc.
3. Clean and disinfect reusable electrodes after each use.
WARNING
Turn off the power before disinfection. The mains supply must be switched off if it is used.
1. Switch off the main unit and disconnect it from the power cord.
2. Wipe the exterior surface of the electrocardiograph, patient cable, and reusable electrodes
(suction bulbs of chest electrodes and the clamps of limb electrodes) using a soft cloth
dampened with the disinfectant solution.
3. Wipe off the disinfectant solution with a dry cloth after disinfection if necessary.
4. Dry the electrocardiograph, patient cable, and reusable electrodes for at least 30 minutes in a
ventilated and cool place.
14.4 Care and Maintenance
CAUTION
Operate the cardiograph, charge the battery, and store the battery at a temperature of
40 (104ºF ) or lower. Exposure to higher temperature may reduce battery life, damage
the battery, and degrade overall cardiograph performance.
14.4.1 Recharge and Replacement of Battery
1) Capacity Identification
The battery capacity can be identified according to the battery symbol in the top right corner
of the LCD screen.
-> -> ->
Capacity is from full to empty.
2) Recharge
SE-12 series electrocardiograph is equipped with the recharge control circuit together with
the battery. When the unit is connected to the mains supply, the battery will be recharged
automatically. Then the battery recharging indicator ( ) and the mains supply indicator
( ) will be lit at the same time. During the recharging course, the symbol flashes in
the top right corner of the LCD screen. After the battery is fully recharged, the symbol stops
flashing, and the battery recharging indicator ( ) is black.