SE-12 Series Electrocardiograph User Manual Safety Guidance
- 8 -
2 Ensure that all device components maintaining personal information (other than
removable media) are physically secure (i.e. cannot remove without tools).
3 Ensure that the electrocardiograph is connected only to the device
authorized/approved by EDAN. Users should operate all EDAN deployed and
supported electrocardiographs within EDAN authorized specifications, including
EDAN approved software, software configuration, security configuration, etc.
4 Protect all the passwords to prevent unauthorized changes. Only the manufacturer’s
service personnel are allowed to modify the System Maintenance Setup.
5 Anti-virus measures such as USB device virus scanning should be carried out prior to
using USB flash drive.
6 When building the networking environment: 1) If a wireless router is used, please turn
on the MAC address filtering function of the wireless router and add the MAC address
of the electrocardiograph to the rule list. The wireless router only allows devices in the
rule list to access the wireless network. 2) It is suggested to build a VLAN, assign the
LAN ports where the approved switch port, electrocardiograph and data management
software are into the same VLAN, and isolate it from other VLANs.
7 When the electrocardiograph is returned for maintenance, disposed of, or removed
from the medical institution for other reasons, it is necessary to ensure that all patient
data are removed from the electrocardiograph (Refer to Chapter 9 Managing Files).
8 Please protect the privacy for the information and the data displayed on the screen,
and for the information and the data stored in the electrocardiograph.
9 Web service is disabled by the electrocardiograph.
1.2.5 Preparation and Operation Warnings (for SE-12 Express
Exercise ECG)
WARNING
1. Test the safety stop (mushroom type) and safety stop (cord type) of the treadmill
before using the system.
2. During the exercise test, ensure that such tests are supervised by properly trained
technician who meets competency requirements for exercise test supervision, fully
trained in cardiopulmonary resuscitation, and is supported by a physician skilled in
exercise testing or emergency medicine who is in close proximity for pretest