SE-12 Series Electrocardiograph User Manual Technical Specifications
Appendix 1 Technical Specifications
A1.1 Safety Specifications
IEC 60601-1:2005/A1:2012
EN 60601-1:2006/A1:2013
Comply with:
IEC 60601-1-2:2014
EN 60601-1-2:2015
IEC/EN 60601-2-25
Anti-electric-shock type:
Class І with internal power supply
Anti-electric-shock degree:
CF type with defibrillation-proof
Degree of
protection against
Ordinary equipment (Sealed equipment without liquid proof)
harmful ingress of water:
Disinfection/sterilization
Refer to the user manual for details
method:
Degree
of
safety
of
application in the presence of
Equipment not suitable for use in the presence of flammable gas
flammable gas:
Working mode:
Continuous operation
EMC:
CISPR 11, Group 1, Class A
Patient
NC
<10μA (AC) / <10μA (DC)
Leakage
SFC
<50μA (AC) / <50μA (DC)
Current:
Patient
NC
<10μA (AC) / <10μA (DC)
Auxiliary
SFC
<50μA (AC) / <50μA (DC)
Current: