SE-12 Series Electrocardiograph User Manual
Operation Preparations
C4
V4
Left mid-axilary line on the same horizontal level as C3
C5
V5
Center of spine on the same horizontal level as C3 and C4
C6
V6
Neck, avoid carotid artery and jugular vein
L
LA
Left deltoid
R
RA
Right deltoid
F
LL
Upper leg as close to torso as possible
N
RL
Upper leg as close to torso as possible
3.3.2 Electrode Placement (for Exercise ECG)
The Precordial Electrodes' Positions on Body Surface:
IEC
AHA
Electrode Placement
C1
V1
Fourth intercostal space at the right border of the sternum
C2
V2
Fourth intercostal space at the left border of the sternum
C3
V3
Fifth rib between C2 and C4
C4
V4
Fifth intercostal space on the left midclavicular line
C5
V5
Left anterior axillary line at the horizontal level of C4
C6
V6
Left midaxillary line at the horizontal level of C4
The Extremity Electrodes' Positions on Body Surface:
IEC
AHA
Electrode Placement
R / L
RA / LA
Below the right/left clavicle
N / F
RL / LL
Below the right/left rib
- 27 -