SE-12 Series Electrocardiograph User ManualOperation Preparations
3.3.3 Attaching the Reusable Electrodes (for Resting ECG)
Connecting to a Lead Wire
Reed
Clamp
Limb Electrode
Suction Bulb
Metal Cup
Connecting to a Lead Wire
Chest Electrode
1) Ensure that the electrodes are clean;
2) Clean the electrode area which is a short distance above the ankle or the wrist with 75%
alcohol;
3) Daub the electrode area on the limb with gel evenly;
4) Place a small amount of gel on the metal part of the limb electrode clamp or on the brim
of the chest electrode's metal cup;
5) Attach the electrodes in place.
When connecting the chest electrodes, place the electrode on the chest electrode site and
squeeze the suction bulb. Unclench it and the electrode is adsorbed on the chest;
NOTE: Long-time measurement with a strong negative pressure on the suction
bulb may cause reddening of the skin. When using the electrode on kids
or patients with delicate skin, squeeze the suction bulb lightly.
3.3.4 Attaching the Disposable Electrodes
CAUTION
The disposable electrodes can only be used for one time.
- 28 -