EasyManua.ls Logo

EDAN SE-1200 Express - Page 81

EDAN SE-1200 Express
115 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SE-12 Series Electrocardiograph User Manual
Operation Instructions for Exercise ECG
Manual Report
Print
Print
Pretest Report
Print
Print
Exercise Report: Auto Report
Print
Print
Exercise Report: Print Time
Late Stage
Late Stage
Recovery Report: Auto Report
Print
Print
Recovery Report: Start Time
20s
20s
Recovery Report: Interval
2min
2min
Final Report: Summary / ST
Scope / Trend Graph / Summary
Print
Print
Template Report
Final Report: Edit Conclusion
Off
Off
Transmission Setup
Items
Default
Default (Only in the U.S.)
Transmit After Saving
Off
Off
FTP User Name/FTP Password
EDANDAT
EDANDAT
FTP Path
Cleared
Cleared
Transmission - WIFI Setup
(with WIFI configured)
Items
Default
Default (Only in the U.S.)
Enable WIFI
Disabled
Disabled
Auto Get IP
Off
Off
Lead Setup
Items
Default
Default (Only in the U.S.)
Lead Sequence
Standard
Standard
Rhythm Lead 1
II
II
Rhythm Lead 2
V1
V1
Rhythm Lead 3
V5
V5
NOTE:
1. Pressing the Up or Down arrow can switch the display style during the exercise test.
When the display style is 3×1, pressing the Left or Right arrow can switch the lead
group.
2. Except for the General Information Setup window, the Record Information Setup
window and the Protocol Manager screen, the changes of the System Setup
screen are synchronous in both Resting and Exercise ECG tests.
- 68 -

Table of Contents

Other manuals for EDAN SE-1200 Express

Related product manuals