SE-12 Series Electrocardiograph User Manual
Hint Information
Message
Cause
U Disk / USB Printer / USB
A U disk, a USB printer or a bar code reader / security reader / ID
Scanner / Reader
card reader is connected to the USB interface.
Table 12-2 Hint Information and Causes for Exercise ECG
Message
Cause
Battery Weak
The battery is weak.
No Paper
Recorder paper runs out or is not loaded.
When Paper Marker is set to Yes, the electrocardiograph
Paper Error
advances the recorder paper to the next black
marker. If it
advances the paper for 300mm and cannot find the next black
marker, the hint Paper Error is displayed.
Module Error
There is something wrong with the signal sample module.
DEMO
The system is in the demonstration mode.
Lead X off
Electrodes fall off the patient or the patient cable falls off the unit.
Overload
The direct current offset voltage on an electrode is too high.
HR Overrange!
Heart rate exceeds the normal range.
Sys. Overrange!
Systolic blood pressure exceeds the normal range.
Dia. Overrange!
Diastolic blood pressure exceeds the normal range.
Lanyard Off
The safety stop (mushroom type) is revolved and the safety stop
(cord type) is pulled out.
Trans.Comm Err
The transducer and the main control board lines are not connected
well.
Transducer Err
The transducer doesn't work well or is not connected well.
Incline Err
The incline motor doesn't work well or is not connected well.
Great Interference
There is great magnetic interference around the transducer.
No Speed Signal
The speed sensor doesn't work well or is not connected well.