SE-3&SE-300 Series Electrocardiograph User Manual Safety Guidance
- 5 -
WARNING
35. Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC/EN 60601-1 approved to the electrocardiograph.
The operation or use of non-approved equipment or accessories with the
electrocardiograph is not tested or supported, and electrocardiograph operation and
safety are not guaranteed.
36. Do not use the additional multiple portable socket-outlet or extension cord in the
medical electrical system, unless it's specified as part of the system by manufacturer.
And the multiple portable socket-outlets provided with the system shall only be used
for supplying power to equipment which is intended to form part of the system.
37. The electrocardiograph shall not be serviced or maintained while in use with a
patient.
38. The appliance coupler or mains plug is used as isolation means from supply mains.
Position the electrocardiograph in a location where the operator can easily access
the disconnection device.
39. The medical electrical equipment needs to be installed and put into service according
to Appendix 2 EMC Information.
40. The equipment should not be used adjacent to or stacked with other equipment, refer
to the recommended separation distances provided in Appendix 2 EMC Information.
41. Portable and mobile RF communications equipment can affect medical electrical
equipment, refer to the recommended separation distances provided in Appendix 2
EMC Information.
42. Assembly of the electrocardiograph and modifications during actual service life shall
be evaluated based on the requirements of IEC60601-1.
1.2.2 Lithium Battery Care Warnings
WARNING
1. Improper operation may cause the lithium battery (hereinafter called battery) to be
hot, ignited or exploded, and it may lead to the declination of the battery capacity. It is
necessary to read the user manual carefully and pay more attention to warning
messages.