SE-3&SE-300 Series Electrocardiograph User Manual Safety Guidance
- 7 -
CAUTION
4. Make sure that there is no intense electromagnetic interference source around the
equipment, such as radio transmitters, mobile phones etc. Attention: large medical
electrical equipment such as electrosurgical equipment, radiological equipment and
magnetic resonance imaging equipment is likely to bring electromagnetic
interference.
5. Before use, the equipment, the patient cable and electrodes etc. should be checked.
Replacement should be taken if there is any evident defectiveness or aging symptom
which may impair the safety or the performance.
6. Ruptured fuse must only be replaced with that of the same type and rating as the
original.
1.2.4 Cleaning & Disinfection Cautions
CAUTION
1. Turn off the power before cleaning and disinfection. If the mains supply is used, the
power cord should be dragged out of the outlet. Prevent the detergent from seeping
into the equipment during cleaning.
2. Do not immerse the unit or the patient cable into liquid under any circumstances.
3. Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
4. Any remainder of detergent should be removed from the unit and the patient cable
after cleaning.
5. Do not use chloric disinfectant such as chloride, sodium hypochlorite etc.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN SE-3 Series and is the answer not in the manual?

EDAN SE-3 Series Specifications

General IconGeneral
Paper TypeThermal paper
Display3.5 inch LCD
Leads12-lead
RecordingThermal printing system
ConnectivityUSB
Power SupplyAC/Rechargeable Battery
Recording ModeAutomatic, Manual, Rhythm
FiltersAC, EMG, and Drift
Paper Size80 mm
BatteryRechargeable Li-ion battery