SE-3&SE-300 Series Electrocardiograph User Manual Safety Guidance
- 2 -
1.2.1 Safety Warnings
WARNING
1. The electrocardiograph is provided for the use of qualified physicians or personnel
professionally trained. They should be familiar with the contents of this user manual
before operation.
2. Only qualified service engineers can install this equipment, and only service
engineers authorized by the manufacturer can open the shell. Otherwise, safety
hazards may happen.
3. Only qualified installation or service engineers can shift the mains supply shift switch
(100V-115V~/220V-240V~) according to local mains supply specifications.
4. The results given by the equipment should be examined based on the overall clinical
condition of the patient, and it can not substitute for regular checking.
5. This device is not intended for treatment or monitoring.
6. The EQUIPMENT is protected against malfunction caused by electrosurgery
according to the clause 36.202.101 in the IEC60601-2-25.
7. Electrodes of dissimilar metals should not be used; it may cause a high polarization
voltage.
8. EXPLOSION HAZARD - Do not use the electrocardiograph in the presence of
flammable anesthetic mixture with oxygen or other flammable agents.
9. SHOCK HAZARD - The power receptacle must be a hospital grade grounded outlet.
Never try to adapt the three-prong plug to fit a two-slot outlet.
10. If the integrity of the external protective conductor is in doubt, the equipment should
be operated by using the built-in rechargeable battery.
11. Do not use this equipment in the presence of high static electricity or high voltage
equipment which may generate sparks.
12. This equipment is not designed for direct cardiac application.
13. Only the patient cable and other accessories supplied by the manufacturer can be
used. Or else, the performance and electric shock protection can not be guaranteed.
The electrocardiograph has been safety tested with the recommended accessories,
peripherals, and leads, and no hazard is found when the electrocardiograph is
operated with cardiac pacemakers or other stimulators.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN SE-3 Series and is the answer not in the manual?

EDAN SE-3 Series Specifications

General IconGeneral
Paper TypeThermal paper
Display3.5 inch LCD
Leads12-lead
RecordingThermal printing system
ConnectivityUSB
Power SupplyAC/Rechargeable Battery
Recording ModeAutomatic, Manual, Rhythm
FiltersAC, EMG, and Drift
Paper Size80 mm