SE-3&SE-300 Series Electrocardiograph User Manual EMC Information
- 78 -
Appendix 2 EMC Information
Guidance and manufacture’s declaration - electromagnetic emissions-
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The Electrocardiograph is intended for use in the electromagnetic environment specified
below. The user of the Electrocardiograph should assure that it is used in such an environment.
Electromagnetic environment – guidance
The Electrocardiograph uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The Electrocardiograph is suitable for use in
all establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies building used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacture’s declaration - electromagnetic immunity -
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The Electrocardiograph is intended for use in the electromagnetic environment specified
below. The user of Electrocardiograph should assure that it is used in such an environment.
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
It is recommended the use
of antistatic materials. If
floor are covered with
synthetic material, the
relative humidity should
be at least 50%.
Electrical fast
transient/burst
2 kV for power
supply lines
2 kV for power
supply lines
It is recommended the use
of filters on power input