SE-3&SE-300 Series Electrocardiograph User Manual Technical Specifications
- 77 -
DC Offset Voltage:
±500mV / ±600mV(optional, with DE12 ECG board)
Minimum Amplitude:
20 μVp-p
Noise:
12.5μVp-p
Multi-channel Crosstalk
0.5mm
Filter
AC Filter: On / Off
DFT Filter: 0.05Hz / 0.15Hz / 0.25Hz / 0.32Hz / 0.5Hz /
0.67Hz
EMG Filter: 25Hz / 35Hz / 45Hz / OFF
LOWPASS Filter:150Hz / 100Hz / 75Hz
CMRR
≥110dB / ≥115dB (optional, with DE12 ECG board)
Sampling Frequency
1000Hz
Pacemaker Detection (Optional, with DE12 ECG board)
Amplitude
±2mV ~ ±700mV
Width
0.1ms ~ 2.0ms
Sampling Frequency
10,000/sec/channel
External Input/Output
Input
100kΩ; Sensitivity 10mm/V±5%;
Single ended
Output
100Ω; Sensitivity 1V/mV±5%;
Single ended
NOTE:
1 Operation of the equipment below the minimum amplitude may cause inaccurate
results.
2 The DE12 ECG board is not available in the U.S.