SE-601 Series Electrocardiograph Service Manual Electrocardiograph Disassembly
- 54 -
3) Disassembly Procedures of the Recorder Casing
Remove the two tapping screws and the two groups of fixing bars, and then the recorder rotation
axis can be dismantled from the recorder casing, as the following figure shows.
4) Disassembly Procedures of the Display Screen for SE-601A
Remove the four cross recessed tapping screws to separate the LCD Screen and Display Screen
Frame, as shown in the following figure.
LCD Screen
Display Screen Frame
Cross Recessed Tapping Screw
Tapping Screws
Recorder Rotation Axis
Fixing Bar
Recorder Casing