4.7.2 Side inlet-port connecon
A side inlet-port is available, refer to Figure: General view of the pump on page 11. To
use the side inlet-port, use the following procedure.
1. Carefully remove the side panel from the pump: use a suitable at blade
screwdriver or similar tool for this purpose. The panel is located by two nylon
rivets.
2. Unscrew and remove the blanking plug with its O-ring.
3.
Unscrew and remove the adaptor ange and its O-ring and replace it with the
blanking plug and O-ring that was removed in Step 2.
4.
Screw the adaptor ange and its O-ring into the side inlet-port.
4.8 Pump outlet connecon
WARNING:
Connect the pump outlet to a suitable treatment plant to prevent the discharge of
dangerous gases and vapours to the surrounding atmosphere. Use a catchpot to
prevent the drainage of contaminated condensate back into the pump.
To connect the pump to your outlet accessories or to the exhaust treatment plant:
Connect a 10 mm internal diameter vacuum hose or 12 mm internal diameter
plasc hose to the outlet nozzle (Figure: General view of the pump on page 11).
Remove the outlet nozzle and connect to the 3/8 inch BSP tapped hole.
Remove the outlet nozzle and replace it with an NW25 ange adaptor (available as
an oponal accessory, see Spares and accessories on page 36) and then connect to
the NW25 ange.
Take note of the following informaon before connecng to the pump outlet:
The exhaust system must be congured so that the maximum pressure at the
pump outlet does not exceed 0.5 bar gauge (1.5 bar absolute, 1.5 x 10
5
Pa) at full
pump throughput.
In the following circumstances, we recommend you to ng an oil mist lter to
the pump outlet
if using the pump with the gas-ballast control open
if operang the pump with an inlet pressure greater than 10 mbar for
extended periods or
if frequently pumping down from atmospheric pressure.
The mist lter will trap the oil exhausted from the pump: the oil can be re-used if it
is not contaminated.
4.9 Leak-test the system
Leaktest the system and seal any leaks found aer you have installed the pump, to
prevent leakage of substances out of the system and leakage of air into the system.
Page 21
A37132880_F - Installaon

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards E2M1.5 and is the answer not in the manual?

Edwards E2M1.5 Specifications

General IconGeneral
BrandEdwards
ModelE2M1.5
CategoryWater Pump
LanguageEnglish

Related product manuals