Table A16 - Inverter alarm codes
INVERTER ALARMS
BIT
set
Fault code
combination
PDT reported fault name Description
1 8 ACCELERATION_TO If motor has not accelerated to a minimum speed of 10Hz
within 60 secs then drive will try to restart 3 times before
giving alarm
4 OVERLOAD_TO Alarm given when dry pump speed is below 7Hz for 3
minutes
2SC_MODE_INTERLOCK Internal fault contact Edwards
1 FLASH_DOWNLOAD_ FAULT Internal fault contact Edwards
2 8 POST_FAULT Internal fault contact Edwards
4 OSTEST _FAULT Internal fault contact Edwards
2 EEPROM_FAULT Internal fault contact Edwards
1 PWM_TRIP Inverter output switched off, active when drive not
running. Restart pump or cycle power to reset.
3 8 MISSING_PHASE_ TIMEOUT Indicates a missing input phase - check power connections
to pump and fuses in supply.
Warning should appear first and then dry pump trips after
10 minutes.
4 EMS EMS has been activated.
Check EMS button and EMS plug in the back of pump.
2 DESAT_FAULT Internal fault contact Edwards
1 UNDERT Inverter is too cold. See warning register to determine
source of problem.