Appendix A2
A591-40-880 Issue A
4 8 OVERT Motor/Drive system over temperature - refer to warning
code to determine source. Check cooling.
4 OVERI Motor overcurrent
2 OVERV Inverter overvoltage
1 Reserved Reserved
Table A17 - Inverter warnings codes
Inverter warnings
BIT
Fault code
combination
Name Description
5 8 Reserved Reserved
4 Reserved Reserved
2 LOWTW Condensation warning, indicates that the inverter (water)
temperature is lower than expected – possible causes
include inverter water valve stuck open.
1 HIGHTW Indicates that the inverter temperature is higher than
expected – possible causes include no or low cooling water
flow, inverter water valve stuck closed or water pipe to
inverter blocked.
6 8 HIGHTC Controller temperature above upper limit / Controller
temperature sensor open circuit - check water supply
4 HIGHTS Inverter heat-sink temperature above upper limit /
Inverter heat-sink temperature sensor open circuit - check
water supply
2 HIGHTM Motor temperature above upper limit / Motor temperature
sensor open circuit - check water supply
1 MISSING_PHASE The loss of one of the three input phases has been
detected - check mains supply to drive or blown fuses
7 8 UTCREG Upper controller temperature regulator active - current
limit reduced - check cooling
4 UTSREG Upper heat sink temperature regulator active - current
limit reduced - check cooling
2 Reserved Reserved
1 LVREG Lower voltage regulator active - check mains voltage
8 8 LOWVCC Internal power supply fault - contact Edwards
4 LOWTC Low controller temperature - contact Edwards
2 LOWTS Low heatsink temperature - contact Edwards
1CAN LOSS Reserved
Table A16 - Inverter alarm codes (continued)
INVERTER ALARMS
BIT
set
Fault code
combination
PDT reported fault name Description