Table 2
Operang and storage condions
Parameter
Units Value
Ambient operang temperature
range
°C 5 to +40
Ambient storage temperature
range
°C -45 to +55
Maximum relave humidity
- 80% for temperatures up to 31 °C de-
creasing linearly to 50% relave hu-
midity at 40 °C
Maximum operang
altude m 2000
Enclosure prot
econ when in-
stalled (IEC60529)
- IP11D
Polluon degree (EN61010) - 2
In
tended use - Indoor
3.3 Manufacturing materials
Table 3
Materials in contact with pumped gases
Material
Where used
Cast iron
Pump, sha and rotorsSteel
Coated cast iron (Xcede variant only)
PTFE
Seals and gas system
Fluoroelastomer
Stainless steel
Gas system
Aluminium
10/2020 - ©Edwards Limited