Characteriscs
Units 3000
1200
1200T
1200TX
600
600T
200
200T
Rated input supply current maximum load
(200 V a.c.)
A (rms) 26 26 26 18
Rated input supply current maximum load
(400 V a.c.)
A (rms) 14 14 14 11
Recommended branch circuit prot
econ UL
(200 230 V a.c. supply)
A 35 35 35 25
Recommended branch circuit protecon UL
(380 460 V a.c.)
A 20 20 20 15
Recommended mains conductor size (mini-
mum)
mm
2
(Awg)
6 (10) 6 (10) 6 (10) 6 (10)
Mains c
onnector - Han® K
4/4-F
Han® 40A
Note:
Inst
all to local, naonal, and regional electrical requirements/codes.
Table 8
General electrical data
Descripon
Units
Rang
Supply voltage 3-phase V a.c. 200 - 230 or 380 - 460 (see rang plate)
Frequency range Hz 50/60
Wiring conguraon 3 wire plus earth (ground)
Branch circuit protecon requirement
Current rang, refer to Table: Pump elec-
trical data
Fuse Class gG (IEC 60269), UL class T, class
J or class RK5, Bussmann type JJS or equiv-
alent I2t characterisc rated to 600 V
Voltage tolerance range ± 10%
Installaon category II (IEC 60664)
Input supply voltage unbalance
Should not exceed 2% when assessed over
any one minute period
Short circuit current rang (when installed
with class T or class J fuses)
kA 30
Second protecve earth (
ground) conductor
Must be installed with cross-seconal ar
ea
equal to phase conductor size, up to 16
mm
2
Typical earth leakage*
for 200 - 230 V a.c. systems
mA < 5
10/2020 - ©Edwards Limited
Page 19M56635880_D
M56635880_D - Technical data