5.3
Stacking
The
TCU
40/80
may
be
stacked
two
high.
Be
sure
that
the
mounting
surface
can
safely
support
the
weight
of
the
two
units
(900
pounds,
408
kg).
Be
sure
that
the
casters
of
the
top
unit
are
turned
inwards,
but
are
not
locked,
until
the
top
unit
is
fully
seated.
5.3.1
Tie-bolts
An
optional
tie-bolt
kit
is
available
for
attaching
stacked
units.
See
Table
15
on
page
52
for
ordering
information.
TCU
40/80
inches
(millimeters)
Figure
7-
Stacked
TCU
40/80
systems
5.4
Facilities
The
TCU
40/80
requires
a
water
supply
flow
rate
between
3 to
6
gallons
per
minute
at
a
maximum
pressure
of
100
psig,
and
an
inlet
temperature
range
of
10
°C
to
26
°C
Power
input
is
by
a
fused,
suitable
isolating
electrical
outlet,
208
VAC,
50/60
Hz,
30
amp,
3-phase
delta
(balanced
load),
3-wire
and
earth.
The
required
receptacle
type
is
Hubbell
P/N
L21-30R2810-A,
or
equivalent.