7.8.4
System
check
Component
settings
Motor
overload
setting
(K3)
Current sensing
relay
setting
(CSR)
Thermostat
setting
(TS1)
High
pressure
switch
setting
(PS1)
3.S
amps
25
amps
210°F
(Differential
set
at
10°F)
300
psig.
(Differential
set
at
40
psig)
Power-up
conditions
Power
cord
connected
to
mains power
supply.
All
circuit
breakers
on.
EMO
released.
Drawer
interlock
switch
in
maintenance
position.
Host
connector
J72J2
installed
(pins
5
&
6
shorted).
READY
lamp
on.
Indicator
Relays
LEDs
Power
Reset
PMR
Current
sensing
relay
Condition
All
off
All
off
On
Off
Green
Off
Note:
REMOTE
RTD
is
optional.
LED
will
be
green
if
used.
Reset
button
1.
Press
RESET.
2.
Verify
contactor
Kl
engages.
3.
Verify
LED
indicator
on
PMR
is
green.
\
Indicator
Relays
LEDs
Power
Reset
PMR
Current
sensing
relay
Condition
CR4on
all
others
off
FACILTITY
POWER,
FACILITY
WATER,
CIRCUIT
BREAKERS,
COOLANT
TEMPERATURE,
COOLANT
NORMAL
LEVEL,
green
COMPRESSOR,
COOLANT
FLOW,,
COOLANT
LOW
LEVEL,
REMOTE
RTD
off
On
On
Green
Off
Note:
REMOTE
RTD
is
optional.
LED
will
be
green
if
used.
32
TCU
40/80 Temperature
Control
Unit