If there is a preparation area, retract the gingiva by gingival restriction
cords. And extract the cords just before scanning the preparation.
Before starting the scan, dry the teeth thoroughly.
During the scanning process, adjust the surgical light to keep the light
away from the patient’s mouth to avoid interference with the scanner.
It is recommended to activate during scanning to automatically
remove soft tissue.
It is recommended to activate during scanning, the scanner
automatically adjusts the scanning light intensity, which is conducive to
the scanning accuracy.
Reusable tips received from the manufacturer are NOT sterilized. You
must sterilize them before the first use.
For detailed information on cleaning, disinfection and sterilization, please
refer to the Helios 500 User Manual: cleaning, disinfection and sterilization.
Avoid any liquid from leaking into the air outlet near the tip mount or the
air inlet at the rear of the scanner (see the figure below), otherwise the
scanner may be damaged.
2) To start scanning, place the tip of the scanner on the surface of the tooth to
stabilize the scanner and press the button on the scaner. Wait until a 3D
image appears in the 3D model display screen, and then slowly move it along
the arch at 0-5mm from the teeth.