This means that the load of a known weight intended for calibration is ready to be
applied.
Important Note: Known weight must be at least 80% of the load cells capacity. (e.g
for a 1000 kg capacity system at least 800 kg must be used)
Once the known weight has been applied (lifted), press OK. The display will show
APPLY followed by the max capacity of the system.
Using the arrow keys ▲▼, set the load value you intend to use for the calibration.
If no key is pressed for 8 seconds, the system will exit out of calibration mode and
you will have to start again from the beginning.
Once the load value on the display matches that of the known weight currently
loaded, press OK.
The display will flash WAIT and then OK.
This means the load cell has successfully been adjusted and the display will return to
the standard measuring screen.
Repeat the procedure above for each load cell.
At any time, you may press ESC to exit out of the calibration process. The display will
read OK and then revert back to the main measuring screen.