6. Error Table
Error codes that may appear on the Ron 2501 indicator display
E=Eilon only
S=User
Serviceable
Calibration or EEprom memory error
Calibration Baud Watch Timer
Scrolling Capacity 10% higher than allowed. Recalibrate. Do not load
the system with a weight higher than the test load.
Totalizer sum value overflow.
Reset the Totalizer according to manual. Only applicable if Totalizer was
ordered
Data Logger memory overflow.
Reset according to manual (see Data Logger). Only applicable if Data
Logger was ordered
Data Logger or Eeprom error.
Reset "Data Logger" according to the manual
Data Logger already active. Wait 5 seconds and Reset Data Logger.
Active Real time Clock (RTC) power. Check/Replace battery.
RS232 or SetPoint still active - wait several seconds and try again.
Eeprom ON/OFF Write/Read.
Reset according to manual (see Totalizer).
Tare attempted above capacity.
Zero Setting: Zero value above 30% of capacity. Reset zero value.
(Multi) Zero = Sum, Tare = Sum
Keyboard error. A key has been pressed more than 3 seconds or
problem with connectors. Check connectors.
Check manual (see Tare). Never use Tare if there is no Tare.
Multiplication factor failure (see manual Multi-Load).
Temperature3 > MAX above 80 deg. Celsius
Data logger does not exist. Read manual.
Menu: Time. Real Time Clock received incorrect data. See manual.
Check baudrate according to manual.
Failure in initial callibration