Sapphire Infusion Pump User Manual43
Figure 2.7. PCA Lockbox 500mL: Workflow
3.
Attach the pump to the mini cradle (Figure 2.7, #3).
4. Hang the IV bag with the medication (Figure 2.7, #4) inside the PCA Lockbox.
5. Clamp the administration set, spike the bag, and attach the administration set to
the pump.
6. Place the administration set through the hole on top of the pump (Figure 2.7, #5).
Verify that there are no kinks in the administration set, so the infusion can run
smoothly.
7. Unclamp the administration set. Then, close both locks of the Lockbox and lock it,
using the supplied key (Figure 2.7, #6).