Fundamental Concepts and Operations 78
Enabling Special Features
Depending on the needs of a specific clinical care area or institution, the Sapphire
pump can be enabled to perform special functions. An authorization code is required
to enable/disable these functions.
The following features can be enabled/disabled by users who have an authorization
code of High:
Local configuration made after the Drug Library is loaded, will be
valid until the user selects a CCA or turns the pump Off.
When Resuming an infusion after pump shutdown, local
configurations will remain until the end of the current infusion. For
more information about Drug Library, refer to Chapter 9: Drug
Library on page 234.
When no Drug Library is loaded, all local changes made to any
configuration will remain valid until reconfigured or until the pump is
set to factory defaults.
Feature
Delivery
Mode(s) Description/Notes
Allow delayed
start
All Enables users to start an infusion at a later time. The
user may either define a specific delay period or set the
pump to Standby.
For more information, refer to Using the Set Delay
Feature on page 227.
Allow
PreProgram
All Allows users to start an infusion using predefined
infusion parameters. When this option is enabled, the
PreSet Programs button appears on the pump Start Up
screen.
For more information, refer to Creating and Editing PreSet
Programs on page 224.