Sapphire Infusion Pump User Manual79
Repeat last
infusion
All Allows the users to start infusions using the same
infusion parameters for the same patient. When this
option is enabled, the Repeat Last Infusion button
appears on the pump Start Up screen.
For more information, refer to Repeating Last Infusion on
page 188.
Limit Period PCA
Epidural
(PCEA and
Epidural
Intermittent)
Specifies the time period to which the dose limit type is
applied (during the selected time, the delivered
boluses will be limited by either maximum number, or
by maximum volume).
Prime Reminder All Enables a reminder for the user to prime the
administration set before starting an infusion.
For more information, refer to Automatic Priming Using
the Pump on page 91
Allow loading
dose
PCA
PCEA
(Epidural)
Enables starting a PCA or PCEA infusion with a
programmed initial clinician dose (bolus).
For more information about enabling the feature, refer
to PCA Options Menu on page 218 or Epidural Mode
Options Menu on page 219.
Password
request
Epidural
(PCEA and
Epidural
Intermittent)
A safety feature that requires password entry to make
changes to important parameters.
For more information about enabling the feature, refer
to Epidural Mode Options Menu on page 219.
Advanced Bolus Continuous Allows users to program a bolus by entering rate,
amount and time.
For more information, refer to Administering a Bolus on
page 111.
Feature