4 : 63101-000 (continued)TEST PROCEDURE US NAVY PPI
To calculate the conductivity, divide the reading displayed by one of the factors shown in the table below:
e 2
Based on an area of 12.5cm and a volume of 3ml.
5 TECHNICAL SPECIFICATION
5.1 ELCOMETER 138 BRESLE SALT KITS
Kit Dimensions
393 x 331 x 95mm (15.5 x 13 x 3.7")
Kit Weight
1.4kg (3lb 1oz)
www.elcometer.com
R
en-10
5.2 BRESLE PATCH
Patch Size
5cm x 5cm
Test Area
2
12.5cm
Sample Volume
3ml
e
Surface Density of Salts Reading
e
Conductivity Value (µS/cm)
ISO Salt Mix
IMO PSPC equivalent NaCI
2
mg/m
Divide by 1.2
Divide by 1.1
µg/cm
2
Divide by 0.12
Divide by 0.11