EasyManua.ls Logo

Elcometer 138 - Post-Test Procedures for ISO 8502-6; ISO 8502-9

Elcometer 138
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
c
6 During a suitable period of time , without removing the needle, suck
d
and re-inject the solution at least four times .
8 Analyse the solution using the Elcometer 138 Meter, see Section 3.4
on page en- . Inject the sample directly into the sensor cell. Rinse the 6
sensor cell several times with the solution to be measured before
taking the reading.
7 At the end of the period, extract as much solution as possible and
d
remove the syringe from the patch .
4 See Appendix A on page en-15 for further instructions on using the Elcometer 138
Meters.
1 Record the temperature of the solution.
2 Remove the patch from the surface and clean the surface. If
required, any adhesive residue from the patch left on the test surface
can be removed by wiping with a cloth moistened with a suitable
solvent. Ensure that the solvent will not damage the surface before
use.
3.3 AFTER TEST
3 Rinse all components of the test kit, other than the patch, in fresh,
pure water. The components can then be used again.
3 : 8502-6 / 8502-9 (continued)TEST PROCEDURE ISO ISO
d
During steps 6 and 7, it is essential that no solution is lost. If any solution is lost, the test shall be rejected.
c
On un-pitted blast-cleaned areas, a period of 10 minutes has been found to be satisfactory, though this time should be agreed by the interested parties.
1
2
3
4
5
R
en-5 www.elcometer.com

Related product manuals