R
33
11 FAULT FINDING / ERROR CODES
To cancel any error code press [ENTER] or [Esc]:
Code Explanation
E01 Calibration has been lost.
E02 Cal difference error is too large/small.
E03 A base value with too great a difference to that currently stored.
E04 Cal was attempted with too large a value.
E05 Base was attempted with too large a value.
E06 Number of readings is greater than 999.
E07 Too many readings stored.
E08 The printer is off-line. (There may be a delay of approximately 30 seconds before receiving
this message).
E09 The low limit is greater than the high limit.
E10 The memory is full.