8.2 OPERATING CONDITIONS AND TESTING
Environmental operating conditions:
Environmental temperature range: from -10°C to +50°C.
Relative humidity (R.H.): from 20% to 80%.
Storage temperature:
from -20°C to +60°C.
Condensation:
non-condensing environment.
Insulation resistance:
500 M between voltmeter input connectors short-circuited between each other and instrument frame, between
power socket and instrument frame.
2 M between voltage and current inputs.
Insulation voltage:
Tested to 2000 Vrms at 50 Hz for 60 seconds between voltmeter input connections (including neutral).
Tested to 3000 Vrms for 60 seconds between each connector and the instrument frame.
Construction standards:
Safety : IEC 1010-1, EN 61010-1, 600V cat. III
EMC : EN 50081-1, EN 50082-1, EN55022.
IEC 801-2, ENV50140 IEC 801-3, IEC 801-4
Conformity : CEE 89/336 (EMC)
CEE 73/23 - CEE 93/68 (Low Voltage Directive)
8.3 POWER REQUIREMENTS
External mains power:
230V ± 10% 50/60 Hz or 110V ± 10% 50/60 Hz
Consumption:
4 VA.
Internal battery power:
6V 940mAh N1-Cd battery with five 1,2V 940mAh elements in series.
8.4 PRIMARY MEASUREMENTS
Measuring method:
fixed sampling, A/D conversion.
Sampling frequency:
2,5 KHz.
Number of samples per phase:
250 (100 msec).
Measuring frequency:
~ 1,2 sec.
Automatic zero regulation:
every 1 minute.