”MANTRIG MAXTIME”)
1 = Continuous: continuous recording of blocks of default 53 seconds duration (interrupt for
storage).
2 = Crest Adv: automatic triggering with Crest factor, enhanced call detection and reduce
noise sensitivity (manual triggering possible to override automatism)
3 = Crest: automatic triggering with Crest factor (manual triggering possible to override
automatism)
4 = Period: automatic triggering with Period algorithm (manual triggering possible to
override automatism)
TRIG_AUTOREC 0 = Manual: A trigger event does not start recording. The recording must be started and
stopped manually, but the trigger frequency is displayed.
1 = Auto: A trigger event automatically starts the recording and recording automatically
stops when trigger conditions are no longer met.
1
TRIG_MODE:
Crest Adv
Min. Crest Minimal Crest factor as trigger condition
Lower values trigger easier
7
Min. F Minimal frequency for trigger condition in kHz 15
Max. F Maximal frequency for trigger condition in kHz 155
TRIG_MODE:
Crest
Min. Crest Minimal Crest factor as trigger condition
Lower values trigger easier
6
Min. RMS Factor for minimal effective value as trigger condition
Lower values trigger easier
2
Min. Peak Factor for minimal amplitude as trigger condition
Lower values trigger easier
2
HP Factor for the lowest frequency, that still may elicit a trigger event (Factor times 1,22 kHz) 6
TRIG_MODE:
Period
ProdVal Period-Trigger: ProdVal
Higher values improve triggering at low frequencies, but also insects, and vice versa.
8
DivVal Period-Trigger: DivVal
Higher values improve triggering at low frequencies, but also insects, and vice versa.
20
Setup 3/7
Audio
Audio settings
Volume Headphones/Speaker
A
-Volume
0 = mute
10 = very loud
ATTENTION: Too high volume can cause hearing loss, particularly using the headphones
5
BATLOGGER Handbuch Elekon AG Nov. 2012 17/28