7.2 Sanitization
Micra should be sanitized once a year. This should be done when the Purication Cartridge is
nearing exhaustion.
During the sanitization cycle a CS1 disinfection tablet is used. A safety data sheet conforming to COSHH
regulations is provided with the tablets and should be read before the tablet is used.
1) TURN system off.
2) UNSCREW carbon lter housing then remove carbon lter.
3) ADD CS1 tablets to housing then allow to dissolve.
4) REPLACE the lter housing (Do not install a new carbon lter).
5) DIRECT Puried Water outlet to Drain.
6) TURN on system for 1 hour.
7) TURN off system. EXCHANGE the Filters and the Purication Cartridge. Refer to page 7
for DI pack installation.
8) RECONNECT Puried Water outlet. System is ready for use.