ELGA Micra Operators Manual
Page 8Micra MANU39752 Version 01 – 02/12
4.7 Reservoir connection
Connect puried water outlet to a suitable reservoir or direct to your nal application.
1) TURN system on – Press the Process button, indicated by arrow, to start. The system is ON when the
green light appears.
LEAVE the system working for about 1 hour to rinse the system. Once completed the system is ready
to use.
CONNECT the power
4.5 Water Connections
Follow instruction, below, when connecting Portable
water inlet, Puried water outlet and Drain outlet.
- UNSCREW compression nut.
- REMOVE transit tting.
- INSERT 1/4” tubing.
- HAND TIGHTEN compression nut.
- 1/2 TURN to extra tighten.
NOTE: At the Portable water inlet, ensure pressure
will not exceed 2 bar (30 psi).
4.6 Commissioning Rinse
Direct both Puried water outlet and Drain outlet to drain before commencing.
Outlet Puried Water
Drain
Potable Water