ELGA Micra Operators Manual
Page 8Micra MANU39752 Version 01 – 02/12
4.7 Reservoir connection
Connect puried water outlet to a suitable reservoir or direct to your nal application.
1) TURN system on – Press the Process button, indicated by arrow, to start. The system is ON when the
green light appears.
LEAVE the system working for about 1 hour to rinse the system. Once completed the system is ready
to use.
CONNECT the power
4.5 Water Connections
Follow instruction, below, when connecting Portable
water inlet, Puried water outlet and Drain outlet.
- UNSCREW compression nut.
- REMOVE transit tting.
- INSERT 1/4” tubing.
- HAND TIGHTEN compression nut.
- 1/2 TURN to extra tighten.
NOTE: At the Portable water inlet, ensure pressure
will not exceed 2 bar (30 psi).
4.6 Commissioning Rinse
Direct both Puried water outlet and Drain outlet to drain before commencing.
Outlet Puried Water
Drain
Potable Water