Model 2000 Flow Computer Instruction Manual
2.0 GENERAL DESCRIPTION
Model 2000 issue 21 Page No
13
2.2.15. MAIN MENU FUNCTION Settings
This menu item is a read only display and when selected will show all the pages of Setting Information available in this
machine these will include:-
General sub-menu items that relate to the machine.
Serial Number up to 8 digits
Contract Hour 0 = Midnight
Energy Units MJ, Kwh, kcal, therms or BTU
Maintenance Mode On or Off
Chromatograph sub-menu items that relate to the gas chromatograph.
Chromatograph type Selected from a list
Communication ID Modbus id
Read Time How often data is read
Chromat status Ignore GC alarms
C6+ code % of higher components
Stream N sub-menu items that relate to the streams
Compressibility Equation in use
Combustion / Base Temperatures
Meter Correction Equations in use
Orifice Equations and calculation methods in use
Energy calculation using Superior or Inferior Heating Values
Total or Flow rate suspension under Low Flow conditions
Total or Flow rate suspension under Alarm conditions
Alarm Counter functions
Gas data alarm use Last Good Value or Keypad
Gas Chromatograph stream number
Station sub-menu items that relate to station functions. Only in a Multistream machine.
Sum Stream into Station Totals or Flow rates
The items will be divided into pages with a maximum of four items per page, each item will be displayed with the
corresponding symbols, scaling factors and units.
2.2.16. MAIN MENU FUNCTION Preset Data
This menu item is a read only display and when selected will show all the pages of Pre-set Data available in this machine
these will include:-
Alarm Levels (not gas data)
All Constants
Base Conditions
Number of Pressure Sensors and On/Off for each
Pressure Sensor selection
Pressure sensor deviation and keypad values
Pressure Units and Gauge or Absolute
Number of Temperature Sensors and On/Off for each
Temperature Sensor selection
Temperature sensor deviation and keypad values
The items will be divided into pages with a maximum of four items per page, each item will be displayed with the
corresponding symbols, scaling factors and units. Any item that has no value or has not been set will be shown as --------