31
Blood Glucose Monitoring System
Control Solution Testing
The purpose of the control solution testing is to make sure the Embrace
®
WAVE+ Blood
Glucose Meter and the Embrace
®
WAVE Blood Glucose Test Strips are working properly.
You should perform control solution testing when:
Using the meter for the rst time
Using a new bottle of Embrace
®
WAVE Blood Glucose Test Strips
You drop the meter
You suspect your meter and test strips are not working properly
The blood glucose test results do not reect how you feel
You want to practice the testing procedure
NOTE:
To test your meter and Embrace
®
WAVE Blood Glucose Test Strips only use the
Embrace
®
WAVE Glucose Control Solution (provided separately). Other brands of control
solution will produce an inaccurate result.