ES120 OPERATION MANUAL
IU ES120 2009-03-01 ENG.3.0
8. System operation and the performance of the system
8.1. Switching on
The system is switched on using a mains switch 1
(Fig. 3).
The accessories can also be
connected to the system when it is on, obviously while keeping appropriate safety conditions.
At this point, care should be taken to prevent a possible switch-on of the generator by
accidental pressing of the push button or pedal.
8.2. Neutral electrode control
When the neutral electrode is not connected or there is a wrong application of the electrode
with an adherence control, after turning on the generator a red indicator lamp flashes and there
is a discontinuous sound signal and a statement displayed: “EL NEU – [] …” activation of the
device in the monopolar mode without an appropriate connection of the neutral electrode is
not possible.
14