100-214-251 REV. 05 59
** The default Start setting is “Standard”.
Auto Seek – Checks horn frequency once each
minute, timed from the last activation of
ultrasonics.
2 – Down (Off)
2 – Up (On)
Auto Seek Duration – Indicates the length of
time the Auto Seek function is active.
3 – Down (100 ms)
3 – Up (500 ms)
Store at End of Weld – Updates horn frequency
memory at the end of each weld.
4 – Down (Off)
4 – Up (On)
Amplitude
Control
Set for Fixed 100% Amplitude, Variable turned
Off.
6 – Up (On)
Start
Short – Sets ramp time to 10 ms.
7 – Down
8 – Down
Medium – Sets ramp time to 35 ms.
7 – Up
8 – Down
Standard** – Sets ramp time to 80 ms.
7 – Down
8 – Up
Long – Sets ramp time to 105 ms.
7 – Up
8 – Up
Table 4.6 Options for Setting the Type 2 Switch Select Block
Functions Options Set Toggle
100-214-251 IWoper.book Page 59 Friday, June 16, 2017 4:12 PM