Loading...
MAS2600: Ordering Information Reference Manual
August 2023 RM 6000-01-001-24
13
Measuring Ranges
MAS 2600 -XXX-XX-X/XX
Gauge Transmitter
The MAS 2600 Gauge Transmitter is supplied in 4 ranges. In the programmable amplifier
the measuring range is divided into 8 sub-ranges and operated by means of the range
select switch.
The transmitter output signal can be adjusted to 20 mA within the sub-ranges as of the
table below.
Absolute Transmitter
The MAS 2600 Absolute Transmitter is available in two ranges.
The output signals are calibrated as follows:
Range
no.
Transducer Type
Type 1:
0 - 3.5 mH
2
O
Type 2:
0 - 7.0 mH
2
O
Type 3:
0 - 16.0 mH
2
O
Type 4:
0 - 35.0 mH
2
O
1 0.20 - 0.37 mH
2
O 0.20 - 0.30 mH
2
O 0.50 - 0.75 mH
2
O 1.00 - 1.90 mH
2
O
2 0.37 - 0,54 mH
2
O 0.30 - 0.50 mH
2
O 0.75 - 1.00 mH
2
O 1.90 - 2.80 mH
2
O
3 0,54 - 0.73 mH
2
O 0.50 - 0.70 mH
2
O 1.00 - 1.50 mH
2
O 2.80 - 3.90 mH
2
O
4 0.73 - 1.10 mH
2
O 0.70 - 1.10 mH
2
O 1.50 - 2.50 mH
2
O 3.90 - 5.80 mH
2
O
5 1.10 - 1.92 mH
2
O 1.10 - 1.80 mH
2
O 2.50 - 4.00 mH
2
O 5.80 - 9.90 mH
2
O
6 1.92 - 3.10 mH
2
O 1.80 - 3.00 mH
2
O 4.00 - 6.00 mH
2
O 9.90 - 15.30 mH
2
O
7 3.10 - 3.50 mH
2
O 3.00 - 4.80 mH
2
O 6.00 - 10.00 mH
2
O 15.30 - 25.10
mH
2
O
8 Not Applicable 4.80 - 7.10 mH
2
O 10.00 - 16.00
mH
2
O
25.10 - 35.60
mH
2
O
Transducer Type
4 0.8 - 3.5 bar abs
7 0.8 - 2.0 bar abs

Related product manuals