Integratethemeterwiththecontrolsystem
Example:Cutoffinteraction
Conguration:
mAOutputProcessVariable=MassFlowRate
FrequencyOutputProcessVariable=MassFlowRate
AOCutoff=10g/s
MassFlowCutoff=15g/s
Result:Ifthemassowratedropsbelow15g/s,alloutputsrepresentingmassowwillreport
zeroow.
Example:Cutoffinteraction
Conguration:
mAOutputProcessVariable=MassFlowRate
FrequencyOutputProcessVariable=MassFlowRate
AOCutoff=15g/s
MassFlowCutoff=10g/s
Result:
Ifthemassowratedropsbelow15g/sbutnotbelow10g/s:
ThemAoutputwillreportzeroow.
Thefrequencyoutputwillreporttheactualowrate.
Ifthemassowratedropsbelow10g/s,bothoutputswillreportzeroow.
2.2.4CongureAddedDamping
DisplayNotavailable
ProLinkII
ProLink→Conguration→AnalogOutput→PrimaryOutput→AOAddedDamp
ProLink→Conguration→AnalogOutput→SecondaryOutput→AOAddedDamp
Communicator
5,3,1,4
DetailedSetup→CongOutputs→AnalogOutput1→PVAOAddedDamping
5,3,2,4
DetailedSetup→CongOutputs→AnalogOutput2→SVAOAddedDamping
AddedDampingcontrolstheamountofdampingthatwillbeappliedtothemAoutput.Itaffects
thereportingofmAOutputProcessVariablethroughthemAoutputonly.Itdoesnotaffectthe
reportingofthatprocessvariableviaanyothermethod(e.g.,thefrequencyoutputordigital
communications),orthevalueoftheprocessvariableusedincalculations.
Note
AddedDampingisnotappliedifthemAoutputisxed(forexample,duringlooptesting)orifthe
mAoutputisreportingafault.AddedDampingisappliedwhilesensorsimulationisactive.
OptionsforAddedDamping
WhenyousetthevalueforAddedDamping,thetransmitterautomaticallyroundsthevaluedown
tothenearestvalidvalue.ValidvaluesarelistedinT able2-4.
8
MicroMotionModel2700TransmitterswithIntrinsicallySafeOutputs