Integratethemeterwiththecontrolsystem
Note
AddedDampingvaluesareaffectedbythesettingofUpdateRateand100HzVariable.
Table2-4ValidvaluesforAddedDamping
UpdateRatesettingProcessvariable
Updaterate
ineffectValidvaluesforAddedDamping
NormalAll20Hz0.0,0.1,0.3,0.75,1.6,3.3,6.5,13.5,27.5,
55.0,110,220,440
100Hzvariable(ifassigned
tothemAoutput)
100Hz0.0,0.04,0.12,0.30,0.64,1.32,2.6,5.4,
11.0,22.0,44,88,176,350
100Hzvariable(ifnot
assignedtothemAoutput)
6.25Hz0.0,0.32,0.96,2.40,5.12,10.56,20.8,43.2,
88.0,176.0,352
Special
Allotherprocessvariables6.25Hz0.0,0.32,0.96,2.40,5.12,10.56,20.8,43.2,
88.0,176.0,352
Interactionofdampingparameters
WhenmAOutputProcessVariableissettoaowvariable,density,ortemperature,Added
DampinginteractswithFlowDamping,DensityDamping,orT emperatureDamping.Ifmultiple
dampingparametersareapplicable,theeffectofdampingtheprocessvariableiscalculatedrst,
andtheaddeddampingcalculationisappliedtotheresultofthatcalculation.
Example:Dampinginteraction
Conguration:
FlowDamping=1sec
mAOutputProcessVariable=MassFlowRate
AddedDamping=2sec
Result:AchangeinthemassowratewillbereectedinthemAoutputoveratimeperiodthat
isgreaterthan3seconds.Theexacttimeperiodiscalculatedbythetransmitteraccordingto
internalalgorithmswhicharenotcongurable.
2.2.5ConguremAOutputFaultActionandmAOutputFaultLevel
DisplayNotavailable
ProLinkII
ProLink→Conguration→AnalogOutput→PrimaryOutput→AOFaultAction
ProLink→Conguration→AnalogOutput→PrimaryOutput→AOFaultLevel
ProLink→Conguration→AnalogOutput→SecondaryOutput→AOFaultAction
ProLink→Conguration→AnalogOutput→SecondaryOutput→AOFaultLevel
Communicator
5,3,1,5
DetailedSetup→CongOutputs→AnalogOutput1→AO1FaultSetup
5,3,2,5
DetailedSetup→CongOutputs→AnalogOutput2→AO2FaultSetup
mAOutputFaultActioncontrolsthebehaviorofthemAoutputifthetransmitterencountersan
internalfaultcondition.
Note
IfLastMeasuredValueTimeoutissettoanon-zerovalue,thetransmitterwillnotimplement
thefaultactionuntilthetimeouthaselapsed.
CongurationandUseManualSupplement