4 Set the switches
Set Alarm and Security switch configuration before installation, as
shown in Figure 4-1.
Figure 4-1: Electronics board
Without LCD display With LCD display
A. Alarm
B. Security
• The Alarm switch sets the analog output alarm to high or low.
• Default Alarm is high.
• The Security switch allows ( ) or prevents ( ) any
configuration of the transmitter.
• Default Security is off (
).
To change the switch configuration:
Procedure
1. If the transmitter is installed, secure the loop, and remove
power.
2. Remove the housing cover opposite the field terminal side.
WARNING
Do not remove the instrument cover in explosive atmospheres
when the circuit is live.
3. Slide the Security and Alarm switches into the preferred
position using a small screwdriver.
Quick Start Guide February2024
14 Emerson.com/Rosemount