2. Place the device onto the mounting socket, ensuring that the
studs of the mounting socket pass through the holes on the
flange of the device. Push the device until the waveguide
makes contact with the pipe and check that the Loctite 5990
has spread out evenly.
3. Place a spring followed by a nut onto each of the four studs.
Then, tighten the nuts until they are touching the top of the
springs.
Note
When tightening the nuts, ensure that the flange remains
parallel to the mounting socket on all sides. This will ensure
that the waveguide of the sensor is flush against the
measurement surface.