3. Add a washer and nut to the U-bolt and tighten until hand
tight. Now, torque the nuts on the U-bolts to 3 N-m. This must
be done a ¼ of a turn at a time until all nuts have reached 3
N-m.
Once the installation of the mount on a small pipe has
been completed, proceed to Installing the Rosemount SAM42
Acoustic Particle Monitor onto the mounting socket to install
the device on the mounting socket.
3.6  Installing the Rosemount SAM42 Acoustic Particle
Monitor onto the mounting socket
This section describes the process of installing the SAM42 Acoustic
Particle monitor onto the mounting socket to ensure that it functions
correctly.
Procedure
1. Place some Loctite 5990 onto the tip of the SAM42 Acoustic
Particle Monitor.