Sand monitoring
To be able to achieve this calibration strategy, the use of a sand
injector skid is required. This method requires both background and
sand injection calibration, together with velocity input. Calibration
curves will be adjusted to meet the well properties over a defined
velocity range.
4.3  Background noise calibration of the device
Procedure
1. To complete background noise calibration, it is recommended
that a minimum of three measurements are taken. These
measurements would be made at the minimum expected flow
velocity of the asset, the maximum expected flow velocity
of the asset and the expected median flow velocity of the
asset. Additional measurements can be taken to improve
the calibration with the flow velocities being in the expected
operating range of the asset.
2. From the commissioning home screen, press the Settings
button.
Quick Start Guide May2024
36 Emerson.com/Rosemount